P/N: 507400 Key Specs
Item
MOBILE BLENDER, 15 TO 40 DEG C, 3 TO 30 LPM PRIMARY OUTLET, 0 TO 30 LPM AUXILIARY OUTLET FLOW, +/-3%, 2.25 IN X 4.10 IN X 4.92 IN, 2.29 LB
OEM
Precision Medical, Inc.
Condition
New OEM Original
Warranty
Other (See Notes)
Accuracy/Linearity
± 1% of Full Scale at Constant Temperature, Relative Humidity and Pressure when Calibrated at Full Scale
Auxiliary Outlet Flow Range
0 LPM - 30 LPM
Battery Life
Approximately 2000 Hr
Bleed Flow
At 50 PSI (3.4 bar) 3 LPM or less
Bypass Alarm Activation
18 PSI - 22 PSI
Diverter Fitting
15 MM "T" adapter
FiO2 Accuracy
± 3% of Full Scale
Flow Rating
3 to 30 LPM Primary Outlet, 0 to 30 LPM Auxiliary Outlet
High Alarm De-Activation Setting
Above 100%
High Alarm Range
PM5900 19% - 100% Oxygen
Low Alarm Range
PM5900 18% - 99% Oxygen
Low Battery Indication
Lo Battery message displayed and audible alarm
Maximum Combined Flow
≥ 30 LPM
Measurement Range
0.0% - 100% Oxygen
Number of Auxiliary Outlet Ports
1
Number of Primary Outlet Ports
1
Operating Conditions Altitude
Sea Level to 8000 FT
Operating Conditions Temperature
50°F - 113°F
Operating Temperature Range
59°F - 104°F
Oxygen Monitor Height
5.44 IN
Oxygen Monitor Length
1.72 IN
Oxygen Monitor Width
3.5 IN
Power Requirements
4 AA Alkaline Battery
Primary Outlet Flow Range
3 LPM - 30 LPM
Response Time
90% of Final Value in Less than 12 SEC at 77°F
Sensor Type
Galvanic Oxygen Sensor
Temperature Rating
15 to 40 Deg C
Total Accuracy
± 3% Actual Oxygen Level Over Full Operating Temperature Range