Product Overview

The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

  • NCG Square x Locking Gland connection
  • Push-To-Set technology for accuracy and safety
  • Factory peak vacuum limit setting
  • Baby icon to indicate Low/Limited model
  • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
  • Pneumatic Unilogic Module for quiet, reliable operation
  • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
  • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
  • Intermittent Mode limited to ASTM Standard 8 LPM flow
  • Modular Components – Simplifies service
  • Low cost replacement parts and kits
  • Durable design and construction, few mechanical parts
  • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
  • "Battery Low" indicator on the digital gauge
  • Technical Specs
    P/N: 8705-1251-905 Key Specs
    Item
    ANALOG DIGITAL VACUUM REGULATOR, NCG SQUARE X LOCKING GLAND, 2.8 IN X 6.7 IN X 4.8 IN
    OEM
    Ohio Medical, LLC
    OEM #
    8705-1251-905
    Condition
    New OEM Original
    Returnable
    Yes
    Warranty
    90 Day
    Connection Type
    NCG Square x Locking Gland
    Certifications
    SBASBA

    The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

  • NCG Square x Locking Gland connection
  • Push-To-Set technology for accuracy and safety
  • Factory peak vacuum limit setting
  • Baby icon to indicate Low/Limited model
  • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
  • Pneumatic Unilogic Module for quiet, reliable operation
  • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
  • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
  • Intermittent Mode limited to ASTM Standard 8 LPM flow
  • Modular Components – Simplifies service
  • Low cost replacement parts and kits
  • Durable design and construction, few mechanical parts
  • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
  • "Battery Low" indicator on the digital gauge
    • Product Overview

      The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

    • NCG Square x Locking Gland connection
    • Push-To-Set technology for accuracy and safety
    • Factory peak vacuum limit setting
    • Baby icon to indicate Low/Limited model
    • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
    • Pneumatic Unilogic Module for quiet, reliable operation
    • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
    • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
    • Intermittent Mode limited to ASTM Standard 8 LPM flow
    • Modular Components – Simplifies service
    • Low cost replacement parts and kits
    • Durable design and construction, few mechanical parts
    • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
    • "Battery Low" indicator on the digital gauge
    • Technical Specs
      P/N: 8705-1251-905 Key Specs
      Item
      ANALOG DIGITAL VACUUM REGULATOR, NCG SQUARE X LOCKING GLAND, 2.8 IN X 6.7 IN X 4.8 IN
      OEM
      Ohio Medical, LLC
      OEM #
      8705-1251-905
      Condition
      New OEM Original
      Returnable
      Yes
      Warranty
      90 Day
      Connection Type
      NCG Square x Locking Gland
      Certifications
      SBASBA

      The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

    • NCG Square x Locking Gland connection
    • Push-To-Set technology for accuracy and safety
    • Factory peak vacuum limit setting
    • Baby icon to indicate Low/Limited model
    • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
    • Pneumatic Unilogic Module for quiet, reliable operation
    • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
    • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
    • Intermittent Mode limited to ASTM Standard 8 LPM flow
    • Modular Components – Simplifies service
    • Low cost replacement parts and kits
    • Durable design and construction, few mechanical parts
    • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
    • "Battery Low" indicator on the digital gauge
      • Product Overview

        The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

      • NCG Square x Locking Gland connection
      • Push-To-Set technology for accuracy and safety
      • Factory peak vacuum limit setting
      • Baby icon to indicate Low/Limited model
      • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
      • Pneumatic Unilogic Module for quiet, reliable operation
      • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
      • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
      • Intermittent Mode limited to ASTM Standard 8 LPM flow
      • Modular Components – Simplifies service
      • Low cost replacement parts and kits
      • Durable design and construction, few mechanical parts
      • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
      • "Battery Low" indicator on the digital gauge
      • Technical Specs
        P/N: 8705-1251-905 Key Specs
        Item
        ANALOG DIGITAL VACUUM REGULATOR, NCG SQUARE X LOCKING GLAND, 2.8 IN X 6.7 IN X 4.8 IN
        OEM
        Ohio Medical, LLC
        OEM #
        8705-1251-905
        Condition
        New OEM Original
        Returnable
        Yes
        Warranty
        90 Day
        Connection Type
        NCG Square x Locking Gland
        Certifications
        SBASBA

        The Ohio Medical® Push-To-SetTM 2M Pediatric and Neonatal Intermittent Suction Unit offers dual modes, an intermittent mode for gastric drainage and a continuous mode for tracheal, pharyngeal or surgical suctioning. The Pediatric and Neonatal Intermittent Suction Unit has a peak vacuum limit** feature to prevent adjusting the vacuum above the medium suction range. The Low Intermittent Suction Unit also contains a unique patented Unilogic timing module that provides a reliable, quiet means of cycling on and off during intermittent suctioning.

      • NCG Square x Locking Gland connection
      • Push-To-Set technology for accuracy and safety
      • Factory peak vacuum limit setting
      • Baby icon to indicate Low/Limited model
      • Positive Pressure Safety and Relief Valve – this feature will vent positive pressure and help prevent patient injury and regulator damage in case of accidental connection to pressurized gas
      • Pneumatic Unilogic Module for quiet, reliable operation
      • Fully adjustable timing cycles (3 to 30 seconds), On and Off independently adjustable
      • Factory preset timing cycles set at approximately 15 seconds On, 8 seconds Off
      • Intermittent Mode limited to ASTM Standard 8 LPM flow
      • Modular Components – Simplifies service
      • Low cost replacement parts and kits
      • Durable design and construction, few mechanical parts
      • Now available with a Digital Gauge - Greater durability and accuracy for support of patient safety, ±1% of full scale
      • "Battery Low" indicator on the digital gauge